Pressure ranges to 20,000 psi (1400 bar). 0.015% of reading accuracy standard (0.008% optional). Single or dual piston format. Built-in hand pumps standard. Mounted spirit level with adjustable feet. Units can be trimmed to local gravity FOC.
The P3100 models are available in single or dual piston format. Unit can be supplied in psi, bar, kgf/cm², and MPa. This robust instrument is highly accurate, quick and easy to use. Units feature a built-in priming pump for large volume applications, piston flotation indicators and a high quality screw press for fine pressure control.
Deadweight testers are the high performance pressure standards for pressure measurement. Utilizing the well proven piston-gauge system consisting of a vertically mounted, precision lapped piston and cylinder assembly, accurately calibrated weight masses (Force) are loaded on the piston (Area), which rises freely within its cylinder. These weights balance the upward force created by the pressure within the system. The pressure is measured when placed on a correctly spinning and floating piston. The total pressure measured is the summation of the weights plus the piston weight carrier assembly.
There are three basic variations in the hydraulic offering; single low pressure, single high pressure and dual piston models. Pressure is generated and controlled by means of a high quality screw press located on the front of the tester. A built-in hand pump is included as standard for all hydraulic models to prime the system and accommodate large volume requirements.
The piston/cylinder assembly is the heart of each deadweight tester. They are manufactured from materials that provide stability, durability, and low thermal coefficients and distortion. Our experience and knowledge of piston/cylinder production and calibration ensure the precision and performance required for today’s demanding calibration requirements.
Standard weight masses are series 3 nonmagnetic austenitic stainless steel. Each mass is marked with the serial number of the instrument and the nominal pressure value relative to the high or low-pressure piston, when applicable. Optional fractional weights are stainless steel and/or solution heat treated aluminum.
Gravity varies significantly with geographical location and this variation has a direct effect on the force of the weights and the accuracy of the deadweight tester. Each instrument can be calibrated to local gravity at no extra cost. If unspecified, instruments will be calibrated to Standard Gravity at 980.665 cm/s2.
| SKU | Image | Product name & short description |
|---|---|---|
| FLUKE P5543 |
![]() |
Fluke P5543 Angle Adapter
The angle adapter uses the standard gauge adapters and positions the gauges at 90º. The maximum working pressure of this unit is 10,000 psi (700 bar).
|
| FLUKE P5544 |
![]() |
Fluke P5544 Two Gauge Stand
This adapter mounts directly to the test port of the calibrator and allows for the calibration of two instruments at the same time or the connection of a reference test instrument.
|
| FLUKE P5551 |
![]() |
Fluke P5551 Pointer Remover and Punch
This tool is designed to quickly remove and consistently refit the pointer of a pressure gauge.
|
| FLUKE P5521 |
![]() |
Fluke Calibration P5521 Liquid-to-Liquid Separator
P5521-For pressures to 10,000 psi (700 bar).
|
| FLUKE 5540-5541 |
![]() |
5540 Metric and 5541 BSP Adapter Sets
5540 Metric and 5541 BSP adapter sets: M14 x 1.5, M20 x 1.5, BSP and NPT thread adapters for pressure calibrators.
|










| SKU Code | Price (EUR) | Stock | |
|---|---|---|---|