Temperature calibration accessories

Type S and Type R Thermocouples 5649, 5650

★ 0.0 (0 reviews)
Sold: 0
Views: 0

Type S and R Thermocouples. Platinum and platinum-rhodium alloy. Temperature range of 0 °C to 1450 °C. Uncertainties as good as 0.15 °C over most of that range. Calibrated at four fixed points (Sn, Zn, Al, Ag) and then extrapolated to 1450 °C. Four version available, 20 and 25 inch lengths with and without CJC.

SKU: FLUKE 5649-20
Stock status: Out of stock
Contact us for pricing and availability.
Description

Key features

  • Type S and R Thermocouples.
  • Platinum and platinum-rhodium alloy.
  • Temperature range of 0 °C to 1450 °C.
  • Uncertainties as good as 0.15 °C over most of that range.
  • Calibrated at four fixed points (Sn, Zn, Al, Ag) and then extrapolated to 1450 °C.
  • Four version available, 20 and 25 inch lengths with and without CJC.

The measuring junction of both the 5649 and the 5650 thermocouples is encased in a 0.25-inch (6.35 mm) alumina sheath that can be ordered in lengths of 20 or 25 inches (50.8 or 63.5 cm) to fit the specific requirements of your application. A reference, or “cold,” junction may also be ordered. The reference junction uses a stainless steel sheath and is 8.25 inches long (21 cm) by 0.188 inches in diameter (4.8 mm). The thin diameter minimizes the immersion depth needed, but the extra length ensures you can get all the immersion you like.

Special tin-plated, solid-copper connecting wires with ultra-low EMF properties are used to help retain the integrity of your thermocouple measurement junction where the probe attaches to your micro voltmeter or Hart Black Stack.

Each thermocouple probe comes with an individual calibration with fixed-point standards to reduce uncertainties to 0.12 °C to 0.37 °C from 232 °C to 962 °C and 0.37 °C to 2.5 °C, increasing linearly >962 °C to 1450 °C (extrapolated).

The probe assembly can be easily disassembled for performing your own bare-wire calibrations.

What's in the box

  • 5649-20 thermocouple
  • Model 2609 case

Manuals + resources


Specifications

  • Range: 0 °C to 1450 °C
  • Type R: Platinum 13 % Rhodium vs. platinum
  • Type S: Platinum 10 % Rhodium vs. platinum
  • Uncertainty, 232 to 962 °C: 0.12 °C to 0.37 °C
  • Uncertainty, above 962 °C: 0.37 °C to 2.5 °C
  • Uncertainty note:
    • k=2, 95 %
    • above 962 °C increasing linearly, extrapolated
  • Hot junction sheath:
    • 6.35 mm diameter
    • 508 mm or 635 mm length
  • Reference junction sheath:
    • 4.8 mm dia.
    • 210 mm long
  • Long-term stability:
    • ± 0.5 °C to 1100 °C
    • ± 2.0 °C to 1450 °C
  • Long-term stability note: over one year depending on usage
  • Short-term stability:
    • ± 0.2 °C to 1100 °C
    • ± 0.6 °C to 1450 °C
  • Immersion: At least 152 mm recommended
  • Protective case: Model 2609 case included
  • Termination code X:
    • B: bare wire
    • W: generic copper-to-copper TC connector
    • R: standard Type R and S TC connector
  • Termination note:
    • reference junction models: no R
    • models without reference junction: no W
  • Weight: 1 kg

Product reviews

Based on 0 reviews
0.0
Overall
5★ 0
4★ 0
3★ 0
2★ 0
1★ 0

Be the first to review “Type S and Type R Thermocouples 5649, 5650”

🎁 Write a review to earn 20 reward points — photo +10, video +15.
Accumulate reward points to redeem unique gifts from Minh Khang Vinatools. Log in
Please log in to submit a review.
Recently viewed
No items yet