Temperature calibrators

Fluke 9144 Field Metrology Well

★ 0.0 (0 reviews)
Sold: 0
Views: 0

Lightweight, portable, and fast. Built-in two-channel readout for PRT, RTD, thermocouple, 4-20 mA current. True reference thermometry with accuracy to ±0.01 °C. On-board automation and documentation. Metrology performance in accuracy, stability, uniformity, and loading.

SKU: FLUKE 9144
Stock status: Out of stock
Contact us for pricing and availability.
Description

Key features

  • Lightweight, portable, and fast
  • Built-in two-channel readout for PRT, RTD, thermocouple, 4-20 mA current
  • True reference thermometry with accuracy to ±0.01 °C
  • On-board automation and documentation
  • Metrology performance in accuracy, stability, uniformity, and loading

High-performance field dry-well for industrial temperature calibration

Built-in Features for Large Workloads and Common Applications

The Fluke 9144 Metrology Well is built to handle a wide range of field applications with speed, precision, and convenience. Whether calibrating 4–20 mA transmitters or thermostatic switches, the 9144 provides:

  • Wide temperature range coverage with three model options: –25 °C to 660 °C
  • Compatibility with PRTs, RTDs, thermocouples, and transmitters
  • Optional “Process” version (914X-X-P) with a built-in two-channel thermometer readout , including 24 V loop power for 4–20 mA transmitters

Each process version:

  • Accepts an ITS-90 reference PRT
  • Features built-in readout accuracy from ±0.01 °C to ±0.07 °C
  • Stores individual sensor calibration constants in the probe housing for interchangeable use
  • Includes a second user-selectable channel for 2-, 3-, or 4-wire RTDs, thermocouples, or transmitters

With Field Metrology Wells, you no longer need to carry multiple devices into the field— one instrument handles it all.

Calibrate the Entire Loop, Not Just the Electronics

Traditionally, only the electronics of a temperature transmitter were calibrated—leaving the sensor unchecked. However, studies show that 75% of transmitter system error originates in the sensor. The 9144 enables full-loop calibration for more accurate and reliable results.

Onboard Automation and Documentation

The process versions include non-volatile memory for storing up to 20 calibration tests, each with:

  • Unique alphanumeric ID
  • Block temperature
  • Reference temperature
  • UUT values
  • Error, date, and time

Results can be:

  • Viewed directly on the front panel
  • Exported via included Model 9930 Interface-it software for reports or raw ASCII data
  • Integrated with the 1586A Super-DAQ for automated sensor calibrations

Automated Thermostatic Switch Testing

Field Metrology Wells support both auto and manual test setups:

  • Auto setup requires just the nominal switch temperature and runs a 3-cycle calibration, displaying the dead band result
  • Manual setup lets you fully customize ramp rate, cycle count, and test sequence

Both options are quick, accurate, and easy to use.

Uncompromising Metrology Performance

The 9144 meets the six key criteria established by EA guidelines for high-accuracy temperature sources:

  • Accuracy
  • Stability
  • Axial (vertical) uniformity
  • Radial (well-to-well) uniformity
  • Loading effects
  • Hysteresis

Each unit comes with an IEC-17025 accredited calibration certificate, backed by:

  • High-quality traceable reference PRTs
  • Rigorous uncertainty analysis
  • 4:1 test uncertainty ratio

Display Accuracy:

  • 9144: ±0.35 °C at 420 °C, ±0.5 °C at 660 °C

Stability:

  • 9144: ±0.05 °C at 660 °C
  • 9142: ±0.01 °C over full range
  • 9143: ±0.02 °C at 33 °C, ±0.03 °C at 350 °C

Radial Uniformity: ±0.01 °C

Axial Uniformity: ±0.05 °C at 40 mm (1.6 in)

Designed for the Field

These wells are engineered for real-world environments:

  • Lightweight and portable: under 8.2 kg (18 lb)
  • Compact footprint for ease of transport
  • Fast time to temperature for productivity in the field
  • Environmental compensation (patent pending) guarantees performance across 13 °C to 33 °C , even in unstable surroundings

User-Friendly Operation

Using the 9144 is straightforward:

  • Large, easy-to-read LCD display
  • Intuitive function keys and menu navigation
  • Dedicated “SET PT.” button for quick temperature input
  • Visual and audible stability indicators
  • Preprogrammed calibration routines stored in memory
  • All inputs are accessible from the front panel

The Metrology Advantage

Fluke Metrology Wells are engineered by the same teams who support national labs and define the Kelvin—not by companies that “dabble” in temperature calibration.

Fluke’s deep expertise in metrology ensures:

  • Published, verifiable specs
  • Unmatched temperature performance
  • Trusted results for demanding industrial environments

Small Dry Wells for Big Field Applications

The 9142, 9143, and 9144 Field Metrology Wells deliver industrial-grade features with laboratory-level accuracy. When combined with the process option, these tools become powerful all-in-one calibration instruments.

Temperature range coverage: –25 °C to 660 °C

Applications:

  • Transmitter loop calibrations
  • Comparison calibrations
  • Thermocouple checks

Field Metrology Wells offer the perfect blend of portability, speed, accuracy, and automation—ideal for demanding field calibration tasks.

What's in the box

  • Users manual
  • Technical manual on CD
  • Serial cable
  • Report of calibration
  • Insert removal tool

Manuals + resources


Specifications

  • Temperature range at 23 °C: 50 °C to 660 °C
  • Display accuracy:
    • ±0.35 °C at 50 °C
    • ±0.35 °C at 420 °C
    • ±0.5 °C at 660 °C
  • Stability:
    • ±0.03 °C at 50 °C
    • ±0.04 °C at 420 °C
    • ±0.05 °C at 660 °C
  • Axial uniformity at 40 mm:
    • ±0.05 °C at 50 °C
    • ±0.35 °C at 420 °C
    • ±0.5 °C at 660 °C
  • Radial uniformity:
    • ±0.02 °C at 50 °C
    • ±0.05 °C at 420 °C
    • ±0.10 °C at 660 °C
  • Loading effect:
    • ±0.015 °C at 50 °C
    • ±0.025 °C at 420 °C
    • ±0.035 °C at 660 °C
    • 6.35 mm reference probe and three 6.35 mm probes
  • Hysteresis: 0.1
  • Operating conditions:
    • 0 °C to 50 °C
    • 0% to 90% RH, non-condensing
  • Environmental conditions, all specs except range: 13 °C to 33 °C
  • Immersion well depth: 150 mm
  • Insert OD: 24.4 mm
  • Heating time: 15 min: 50 °C to 660 °C
  • Cooling time:
    • 25 min: 660 °C to 100 °C
    • 35 min: 660 °C to 50 °C
  • Display resolution: 0.01°
  • Display: LCD, °C or °F user-selectable
  • Power requirements:
    • 100 V to 115 V ±10%, 50/60 Hz, 1400 W
    • 230 V ±10%, 50/60 Hz, 1800 W
  • Computer interface: RS-232 interface
  • Calibration: Accredited calibration included
  • Automation: use the 1586A Super-DAQ to automate sensor calibration
  • Reference readout accuracy, 4-wire probe:
    • ±0.010 °C at –25 °C
    • ±0.015 °C at 0 °C
    • ±0.020 °C at 50 °C
    • ±0.025 °C at 150 °C
    • ±0.030 °C at 200 °C
    • ±0.040 °C at 350 °C
    • ±0.050 °C at 420 °C
    • ±0.070 °C at 660 °C
  • Reference readout note:
    • temperature range may be limited by the reference probe
    • probe uncertainty and characterization errors not included
  • Reference resistance range: 0 ohms to 400 ohms
  • Reference resistance accuracy:
    • 0 to 42 ohms: ±0.0025 ohms
    • 42 to 400 ohms: ±60 ppm of reading
  • Reference characterizations: ITS-90, CVD, IEC-751, Resistance
  • Reference measurement capability: 4-wire
  • Reference probe connection: 6-pin DIN with Infocon Technology
  • RTD readout accuracy:
    • NI-120: ±0.015 °C at 0 °C
    • PT-100 385: ±0.02 °C at 0 °C
    • PT-100 3926: ±0.02 °C at 0 °C
    • PT-100 JIS: ±0.02 °C at 0 °C
  • RTD resistance range: 0 ohms to 400 ohms
  • RTD resistance accuracy:
    • 0 to 25 ohms: ±0.002 ohms
    • 25 to 400 ohms: ±80 ppm of reading
  • Accuracy note:
    • within operating range, 4-wire PRTs
    • 3-wire RTDs: add 0.05 ohms plus lead wire resistance difference
  • RTD characterizations:
    • PT-100 385, JIS, 3926
    • NI-120
    • Resistance
  • RTD measurement capability:
    • 4-wire RTD
    • 2-, 3-wire RTD with jumpers only
  • RTD connection: 4 terminal input
  • TC readout accuracy:
    • Type J: ±0.7 °C at 660 °C
    • Type K: ±0.8 °C at 660 °C
    • Type T: ±0.8 °C at 400 °C
    • Type E: ±0.7 °C at 660 °C
    • Type R: ±1.4 °C at 660 °C
    • Type S: ±1.5 °C at 660 °C
    • Type M: ±1.4 °C at 660 °C
    • Type L: ±0.7 °C at 660 °C
    • Type U: ±0.75 °C at 600 °C
    • Type N: ±0.9 °C at 660 °C
    • Type C: ±1.1 °C at 660 °C
  • TC millivolt range: –10 mV to 75 mV
  • Voltage accuracy: 0.025% of reading + 0.01 mV
  • Internal cold junction compensation accuracy: ±0.35 °C, ambient 13 °C to 33 °C
  • TC connection: Small connectors
  • mA readout accuracy: 0.02% of reading + 2 mV
  • mA range: Cal 4-22 mA, Spec 4-24 mA
  • mA connection: 2 terminal input
  • Loop power function: 24 V DC loop power
  • Electronics temperature coefficient:
    • ±0.005% of range per °C
    • at 0-13 °C and 33-50 °C
  • Size H x W x D: 290 x 185 x 295 mm
  • Weight: 7.7 kg
Optional Accessories
SKU Image Product name & short description
FLUKE 9144-INSA Fluke 9144-INSA Insert “A” 9144, Imperial Misc Holes
Fluke 9144-INSA Insert “A” 9144, Imperial Misc Holes
Fluke 9144-INSA Insert “A” 9144, Imperial Misc Holes.
FLUKE 9144-INSB Fluke 9144-INSB Insert “B” 9144, Imperial Comparison Holes
Fluke 9144-INSB Insert “B” 9144, Imperial Comparison Holes
Fluke 9144-INSB Insert “B” 9144, Imperial Comparison Holes.
FLUKE 9144-INSC Fluke 9144-INSC Insert “C” 9144, 0.25-inch Holes
Fluke 9144-INSC Insert “C” 9144, 0.25-inch Holes
Fluke 9144-INSC Insert “C” 9144, 0.25-inch Holes.
FLUKE 9144-INSD Fluke 9144-INSD Insert “D” 9144, Metric Comparison Holes
Fluke 9144-INSD Insert “D” 9144, Metric Comparison Holes
Fluke 9144-INSD Insert “D” 9144, Metric Comparison Holes.
FLUKE 9144-INSE Fluke 9144-INSE Insert “E” 9144, Metric Misc Holes w/0.25-inch Hole
Fluke 9144-INSE Insert “E” 9144, Metric Misc Holes w/0.25-inch Hole
Fluke 9144-INSE Insert “E” 9144, Metric Misc Holes w/0.25-inch Hole.
FLUKE 9144-INSY -
Fluke 9144-INSY Insert Custom 9144
Fluke 9144-INSY Insert Custom 9144.
FLUKE 9144-INSF Fluke 9144-INSF Insert “F” 9144, Metric Comparison Misc Holes w/0.25-inch Hole
Fluke 9144-INSF Insert “F” 9144, Metric Comparison Misc Holes w/0.25-inch Hole
Fluke 9144-INSF Insert “F” 9144, Metric Comparison Misc Holes w/0.25-inch Hole.
FLUKE 9144-INSZ Fluke 9144-INSZ Insert “Z” 9144, Blank
Fluke 9144-INSZ Insert “Z” 9144, Blank
Fluke 9144-INSZ Insert “Z” 9144, Blank.
FLUKE HART-DRYWELL-CABLE HART DRYWELL CABLE Interface Cable Kit
HART DRYWELL CABLE Interface Cable Kit
An interface cable kit that includes a null modem to connect the Fluke 754 with Fluke Calibration dry-wells and Micro-Baths.

Product reviews

Based on 0 reviews
0.0
Overall
5★ 0
4★ 0
3★ 0
2★ 0
1★ 0

Be the first to review “Fluke 9144 Field Metrology Well”

🎁 Write a review to earn 20 reward points — photo +10, video +15.
Accumulate reward points to redeem unique gifts from Minh Khang Vinatools. Log in
Please log in to submit a review.
Recently viewed
No items yet